Metrel BUNDLE S 2115 Safety Test System with Current-Limitation
Limit high-voltage test current. Keep test initiation under control.
Plan an end-of-line test station around a high-voltage test current limit of ≤3 mA. The Metrel BUNDLE S 2115 Safety Test System with Current-Limitation combines the MI 3394 CE MultiTester XS with current monitoring, safety interlocking and external start control.
An emergency stop circuit and monitored outputs support controlled operation, with hazardous outputs disabled when a fault or safety trigger occurs.
Request a Metrel BUNDLE S 2115 quote today. Tell us about your test station requirements.

Metrel BUNDLE S 2115 Key Features:
- Current-limited HV testing: test current restricted to ≤3 mA
- Continuous supervision: monitors output current and safety conditions
- Deliberate test initiation: external start control requires operator action
- Emergency stop protection: shuts down hazardous outputs on intervention or fault
- Automated testing: supports AutoSequence®-compatible test sequencing
Why choose the Metrel BUNDLE S 2115?
Choose this bundle where the HV test method permits a maximum current of 3 mA. Its low-energy approach can simplify guarding requirements in line with EN 50191 principles, subject to correct integration, the station’s risk assessment and applicable national requirements.

What Comes with the Metrel BUNDLE S 2115?
- MI 3394 CE MultiTester XS
- S 2115 safety system with current protection
- (A 1496) Two-LED warning lamp with magnet
- (A 1499) External power supply, 24 V
- (A 1495) PL adaptor for pedal and signal lamps
- (A 1584) Rack mount adaptor
- (A 1690) Start button and A 1689 stop button
- (A 1017) RS232 cable and A 1727 USB Type A/B cable
- (A 1728) Power cable, 2 m, 3 × 0.75 mm²
- (A 1493) Power cable, 2 m, 3 × 1.5 mm²
- (A 1695) Yellow/green earth cable, 2 m, 2.5 mm²
- (A 1602) Blue HV test lead, 1.5 m, 5 kV
- 2 × (A 1773) Red mini HV crocodile clips
- Calibration certificate, short instruction manual and instruction manual
- Metrel ES Manager BASIC licence and SW 1201 software installation
Test cages and protective enclosures are not included.
| Features | |
| Core instrument | MI 3394 CE MultiTester XS |
| Safety system | S 2115 with current limitation and monitoring |
| HV test current limit | Itest ≤3 mA |
| High-voltage testing | Programmed AC/DC withstand tests |
| Protective earth / continuity testing | 4-wire |
| Insulation resistance testing | ✓ |
| Leakage current tests | Substitute, differential, touch and PE |
| Discharge time measurement | ✓ |
| Functional testing | P/S/Q, voltage, current, cos φ, frequency, voltage/current THD and PF |
| Automated sequencing | AutoSequence® compatible |
| Operator controls | External start and emergency stop |
| External power supply | A 1499, 24 V |
| Warning indication | A 1496 two-LED lamp with magnet |
| Software | Metrel ES Manager BASIC; SW 1201 |
| Calibration certificate | ✓ |
❓ Frequently Asked Questions (FAQ)
Q1: What does the 3 mA limit apply to?
A. It applies to the current-limited high-voltage test arrangement. The specified test current is Itest ≤3 mA.
Q2: Can physical guarding be omitted automatically?
A. No. Any reduction in guarding depends on correct implementation, the station’s risk assessment and applicable national requirements.
Q3: Can the system run automated test sequences?
A. Yes. The bundle supports automated test sequencing compatible with AutoSequence®.
Request a Quote for the Metrel BUNDLE S 2115
Send us your HV test requirements, device details and proposed station arrangement. We can help you assess whether the current-limited approach fits your application.