Megger OTS60AF Oil Test System - Portable Liquid Dielectric Breakdown Tester
Megger OTS60 60kV AF models have much a larger test chamber for even easier access and cleaning, particularly useful in a lab environment.  They are fitted with a 12 key alpha-numeric keypad to facilitate entry of test ID, file names, notes etc.  Alpha characters are entered by repetitive pressing on a key.  The AF models also have the ability to use a USB barcode reader to scan oil sample barcode labels, ideal for better integration with a LIM system.
 
▶ Video: Megger OTS Oil Testers Overview

 
Megger OTS Key Features:
- Full Range to suit all user needs
 - Easy adjust – locking electrode gap
 - Fast precision breakdown detection
 - Ultra-fast HV switch off time
 - Suitable for mineral, ester and silicone oils
 - Flat electrode gap gauges that will not damage electrodes
 - Oil temperature is measured continuously so it can be determined whether the oil test sample is within the range allowed by the test standards before the test is commenced
 - QVGA colour display with back-light (easy to read in sunlight or dark conditions)
 - Large, easy clean test chamber with oil drain
 - High visibility test chamber
 - Safe operation with dual redundant micro switches
 - Intuitive user interface
 - Fully automatic operation with pre-loaded international test standards
 - User configurable test sequences to cover transition period of new / updated test standards (standards maintained via USB updates from Megger)
 - All instruments supplied with one 400 ml test vessel in the box as standard, unless the super user kit is specified (see below)
 - Built onto rigid box section chassis to prevent flexing on impact that otherwise would damage transformer
 - Unique built in chamber drain pipe for easy removal of oil accidentally spilt in test chamber, this can easily be connected to a lab waste system
 - Test standards favourites list speeds up selection by only displaying the standards regularly used by the use
 - NEW - IEC60156-2018
 - NEW - IEC60156-2018 V for Viscous Oils
 - NEW - IEC60156-2018 Annex A
 - NEW - IEC60156-2018 Annex A (V) for Viscous Oils
 - NEW - Withstand tests
 
 
Megger OTSAF Extra Key Features:
- Barcode scanning capability for oil sample ID
 - Extra large test chamber for ease of use in high productivity application
 - 12 key alpha-numeric keypad to facilitate entry of test ID, file names, notes etc
 
 
What comes with the Megger OTS60AF?
- OTS60AF 60kV Auto laboratory based oil test set
 - Vessel 400 ml assembly (stirrer lid fitted)
 - Printer
 - Full electrode set - IEC and ASTM
 - Electrode gauge set
 - IEC and ASTM impeller
 - Calibration certificate
 - Power DB guide
 - Quick Start Guide
 - OTS Vessel Preparation Guide
 
                               
            
                Note: Be sure to check the official datasheet before buying, and the user manual before using any test equipment. Need training? Just reach out - we’re happy to help